USING THE COMBISCOPE INSTRUMENTS 3 - 37
The following diagram shows the possible states of the acquisition process when
"averaging" is on, and the way they are affected by commands.
Figure 3.13 The Trigger Model during acquisition averaging
INIT
or
INIT:CONT ON
*RST
ABORt
power on
No
Yes
IDLE state
INITiated state
INIT:CONT ON
No
Yes
Wait for completeWait for trigger
1
acquisition
+
averaging
TRIGger
:LEVel
:SLOPe
TRIGger
:SOURce
LINE
INTernal
IMMediate
ST7189
SENSe:AVERage:COUNt
No
Wait for AVERage state
Yes