USING THE COMBISCOPE INSTRUMENTS 3 - 25
3.4.1.3 The trigger modes
A combination of the INITiate:CONTinuous and TRIGger:SOURce command
allows you to define the following trigger modes:
Table 3.1 The TRIGger modes
Trigger mode:
INITiate
:CONTinuous
TRIGger
:SOURce
>>>Single-shot<<<
Generates one sweep, regardless of any
trigger settings (valid after
*
RST).
OFF IMMediate
>>>Single-shot<<<
Generates one sweep, triggered using
trigger settings.
OFF
INTernal<n>
or
LINE
>>> Single-shot <<<
Generates one sweep, externally triggered
via channel 4 (only for PM33x0B).
OFF EXTernal
>>>Auto trig<<<
Generates continuous sweeps,
independent of any trigger settings.
ON IMMediate
>>>Normal trig<<<
Generates continuous sweeps, triggered
using trigger settings.
ON
INTernal<n>
or
LINE
>>> Normal trig <<<
Generates continuous sweeps, externally
triggered via channel 4 (only for PM33x0B).
ON EXTernal
>>>Single-Shot<<<
Generates one sweep triggered by
*
TRG
or GET, regardless of any trigger settings.
ON
or
OFF
BUS