® 80 Series V Digital Multimeter Calibration Manual PN 2102915 September 2004, Rev.1, 2/05 © 2004-2005 Fluke Corporation, All rights reserved. P
80 Series V Calibration Manual 2 XWWarning To avoid possible electric shock or personal injury, inspect the test leads for damaged insulation or expos
Digital Multimeter Precautions and Safety Information 3 • When making electrical connections, connect the common test lead before connecting t
80 Series V Calibration Manual 4 Electrical Symbols Electrical symbols used on the Meter and in this manual are explained in Table 1. Table 1. Electri
Digital Multimeter Specifications 5 664 to 600 V Overvoltage Category IV. UL listed to UL61010-1. Licensed by TÜV to EN61010-1. Detailed Specif
80 Series V Calibration Manual 6 Table 4. DC Voltage, Resistance, and Conductance Function Specifications Accuracy Function Range Resolution Model 83
Digital Multimeter Specifications 7 Table 7. Capacitance and Diode Function Specifications Function Range Resolution Accuracy E 10.00 nF 100
80 Series V Calibration Manual 8 Table 11. Electrical Characteristics of the Terminals Function Overload Protection1 Input Impedance (nominal) Common
Digital Multimeter Basic Maintenance 9 Basic Maintenance XWWarning To avoid possible electric shock or personal injury: • Remove the test lead
80 Series V Calibration Manual 10 WCaution To avoid damaging the Meter, the gasket that is sealed to the bottom case, and is between the two case halv
Digital Multimeter Basic Maintenance 11 5. Place the PCA onto the top shield so that the screw holes align. 6. Place the bottom shield onto t
Lifetime Limited Warranty Each Fluke 20, 70, 80, 170 and 180 Series DMM will be free from defects in material and workmanship for its lifetime. As
80 Series V Calibration Manual 12 Replacing the Battery Replace the battery with a 9-V battery (NEDA A1604, 6F22, or 006P). XWWarning To avoid false r
Digital Multimeter Required Equipment 13 MIN MAXRANGERELHz %AutoHOLDPeak MIN MAX4½ DIGITS1 Second˚C/˚FOFFmA AmVVVAAmACOMV400mAFUSED10A MAXFUSED
80 Series V Calibration Manual 14 Table 13. Required Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltage Range: 0 -
Digital Multimeter Performance Tests 15 Basic Operability Tests Refer to the following sections to test the basic operability of the Meter. Tes
80 Series V Calibration Manual 16 Table 13. Display Features (cont.) Number Feature Indication L b The battery is low. XWWarning: To avoid false rea
Digital Multimeter Performance Tests 17 Testing the Pushbuttons To test the pushbuttons 1. Turn the Meter rotary knob to J. 2. Press each bu
80 Series V Calibration Manual 18 Table 14. Accuracy Tests (cont.) Display Reading Step Test Function Range 5500A Output 83 and 88 87 23 600 Ω 330
Digital Multimeter Performance Tests 19 Table 14. Accuracy Tests (cont.) Display Reading Step Test Function Range 5500A Output 83 and 88 87
80 Series V Calibration Manual 20 6. Set up the function generator as follows: Frequency 900 kHz Amplitude 8.3 V Burst Mode yes Burst Count 1 Bur
Digital Multimeter Performance Tests 21 INDUCTIVEPICKUPSPARK PLUG SIDEOSCILLOSCOPEFUNCTION GENERATORCH1CH210 RESISTORWIRE LOOP(14 AWG)OUTPUTRED
i Table of Contents Title Page Introduction...
80 Series V Calibration Manual 22 aad09f.wmf Figure 6. Waveform for Inductive Pickup Test
Digital Multimeter Calibration Adjustment 23 Calibration Adjustment The Meter features closed-case calibration adjustment using known reference
80 Series V Calibration Manual 24 Restoring the Default Password If the calibration password is forgotten, the default password (1234) can be restored
Digital Multimeter 25 Meter Buttons Used in the Calibration Steps The Meter buttons behave as follows when performing the Calibration Adjustm
80 Series V Calibration Manual 26 Notes Set the calibrator to Standby prior to changing the function switch position and or after completing adjustmen
Digital Multimeter Service and Parts 27 Service and Parts Replacement parts are shown in Table 17 and Figure 8. To order parts and accessories,
80 Series V Calibration Manual 28 H2-4 (3)BT1MP390-391MP6MP22MP10-11MP5 HolsterMP15 Alligator ClipsTM3TM2TL75Test Lead SetTilt StandMP8TM1WF2F1MP14S2M
80 Series V Calibration Manual ii
iii List of Tables Table Title Page 1. Electrical Symbols ...
80 Series V Calibration Manual iv
v List of Figures Figure Title Page 1. Opening the Meter, Battery and Fuse Replacement... 10 2. R
80 Series V Calibration Manual vi
1 Introduction XW Warning To avoid shock or injury: • Read “Precautions and Safety Information” before performing the verification tests or cali