1625
Users Manual
58
Function Displays Condition Note
E
LIMIT
edw041.eps
Display of
measured
value ex-
ceeds
LIMIT.
Measured
value is
higher than
set LIMIT.
K
LIMIT
edw042.eps
Compen-
sation
higher than
measured
value.
Delete
compensa-
tion or
switch in-
strument
ON/OFF.
ST
edw043.eps
Wrong po-
larity on
jacks E
and ES.
Reverse
polarity.
E
edw044.eps
Measured
value in-
stable.
Unsteady
noise volt-
age. Try
time aver-
age meas-
urement.
ST
V
edw045.eps
Current in
external
trans-
former to
low.
Reduce
auxiliary
current
spike re-
sistance.
Legend: W= displayed flashing