Impulse 6000D/7000DP Users Manual Supplement
8/08 5
Normal Sinus waves
Lead # I II III V1 V2 V3 V4 V5 V6
Ref. Amp. 70 % 100 % 30 % 24 % 48 % 100 % 120 % 112 % 80 %
Lead I reference
Performance waves and R wave detection
Lead # I II III V1 V2 V3 V4 V5 V6
Ref. Amp. 100 % 150 % 50 % 100 % 100 % 100 % 100 % 100 % 100 %
Normal Sinus waves
Lead # I II III V1 V2 V3 V4 V5 V6
Ref. Amp. 100 % 150 % 50 % 24 % 48 % 100 % 120 % 112 % 80 %
ECG Normal Sinus
Rates................................................................... 10 to 360 (by 1) BPM
ECG High Level Output (BNC Jack)
Amplitude
Range ............................................................. 0.5 V per mV of reference lead setting
Accuracy ......................................................... ±5 %
Output Impedance .............................................. 50 Ω
ECG on Defibrillator Input Load
Same as the LEAD II amplitude but limited to ±4 mV
ECG Performance Waves
Square wave ....................................................... 2.0 and 0.125 Hz
Triangular wave .................................................. 2.0 and 2.5 Hz