Fluke IMPULSE 6000D Manual de usuario Pagina 3

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 11
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 2
Manual Supplement Impulse 6000D/7000DP Users
2 8/08
Accuracy............................................................. ±0.1 ms
Voltage
Range ................................................................. 20 to 5000 V
Accuracy............................................................. ±(1 % of reading + 2 V)
Current
Range ................................................................. 0.4 to 100.0 A
Accuracy............................................................. ±(1 % of reading + 0.1 A)
Tilt (biphasic and pulsed biphasic)
Range ................................................................. 1 % to 99 %
Accuracy............................................................. ±1 digit
Interphase delay (biphasic and pulsed biphasic)
Range ................................................................. 0.1 ms to 9.9 ms
Accuracy............................................................. ±0.1 ms
Frequency (pulsed biphasic only)
Range ................................................................. 2000 Hz to 8000 Hz
Accuracy............................................................. ±1 % of reading
Duty cycle (pulse biphasic only)
Range ................................................................. 1 % to 99 %
Accuracy............................................................. ±1 digit
Sample rate ........................................................... 250 kHz (4 μs sample)
Maximum Average Power.................................... 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Vista de pagina 2
1 2 3 4 5 6 7 8 9 10 11

Comentarios a estos manuales

Sin comentarios