Fluke 77 Manual de usuario Pagina 11

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 12
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 10
77/75/23/21 English Instruction Sheet
Page
11
Function Range Accuracy
L
3.200 V, 32.00 V, 320.0 V
600 V (21/75), 1000 V (23/77)
±(0.3%+1)
±(0.4%+1)
m
L
320.0 mV ±(0.3%+1)
K (45 to 500 Hz,
3.2 V range. Other
ranges 45 to 1 kHz)
3.200 V, 32.00 V, 320.0 V
600 V (21/75), 1000 V (23/77)
±(2%+2)
±(2%+2)
21/75:
23/77:
320.0
3200 , 32.00 k,
320.0 k 3.200 M
32.00 M
320.0
3200 , 32.00 k,
320.0 k 3.200 M
32.00 M
±(0.5%+2)
±(0.5%+1)
±(0.5%+1)
±(2%+1)
±(0.5%+3)
±(0.5%+1)
±(0.5%+1)
±(2%+1)
2.0 V ±(1% typical)
Function Range Accuracy Burden
Voltage
(typical)
? (45 Hz
to 1 kHz)
32.00 mA, 320.0 mA
10.00 A *
±(2.5%+2)
±(2.5%+2)
6 mV/mA
50 mV/A
A 32.00 mA, 320.0 mA
10.00 A *
±(1.5%+2)
±(1.5%+2)
6 mV/mA
50 mV/A
* 10 A continuous, 20 A for 30 seconds maximum.