Models 175, 177 & 179 Users Manual 6 Making Basic Measurements The figures on the following pages show how to make basic measurements. When conne
True RMS Multimeters Making Basic Measurements 7 Testing for Continuity RANGEHOLDMIN MAXRANGEHOLDMIN MAX AIK06F.EPS Measuring Temperature (Mod
Models 175, 177 & 179 Users Manual 8 Measuring AC or DC Current XW Warning To avoid personal injury or damage to the Meter: • Never attempt to m
True RMS Multimeters Using the Bar Graph 9 Measuring Frequency XW Warning To avoid electrical shock, disregard the bar graph for frequencies &
Models 175, 177 & 179 Users Manual 10 Cleaning Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Dirt or moi
True RMS Multimeters Specifications 11 Specifications Accuracy is specified for 1 yr after calibration, at operating temperatures of 18 °C to
Models 175, 177 & 179 Users Manual 12 Accuracy ±( [ % of Reading ] + [ Counts ] ) Function Range 1 Resolution Model 175 Model 177 Model 179 AC
True RMS Multimeters Specifications 13 Accuracy ±( [ % of Reading ] + [ Counts ] ) Function Range 1 Resolution Model 175 Model 177 Model 179
Models 175, 177 & 179 Users Manual 14 Function Overload Protection 1 Input Impedance (Nominal) Common Mode Rejection Ratio (1 kΩ Unbalanced) Nor
Lifetime Limited Warranty Each Fluke 20, 70, 80, 170 and 180 Series DMM will be free from defects in material and workmanship for its lifetime. As use
i Table of Contents Title Page Contacting Fluke...
ii XWWarning. Read before using the Meter: To avoid possible electrical shock or personal injury, follow these guidelines: ⇒ Use the Meter only
1 Models 175, 177 & 179 True RMS MultimetersThe Fluke Model 175, Model 177, and Model 179 are battery-powered, true-RMS multimeters (hereafter &q
Models 175, 177 & 179 Users Manual 2 Battery Saver ("Sleep Mode") The Meter enters the "Sleep mode" and blanks the display if
True RMS Multimeters Display 3 Display . 8775643211314159111210AIK02F.EPS No. Symbol Meaning 1 s Continuity test. 2 O Diode test. 3 K Negativ
Models 175, 177 & 179 Users Manual 4 MIN MAX AVG Recording Mode The MIN MAX AVG recording mode captures the minimum and maximum input values, and
True RMS Multimeters Manual Ranging and Autoranging 5 Manual Ranging and Autoranging The Meter has both Manual range and Autorange modes. ⇒ I