Fluke DDS Function Generator 271 Manual de usuario Pagina 7

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 34
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 6
EMC Compliance
This instrument meets the requirements of the EMC Directive 89/336/EEC.
Compliance was demonstrated by meeting the test limits of the following standards:
Emissions
EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits
used were:
a) Radiated: Class B
b) Conducted: Class B
c) Harmonics:EN61000-3-2 (2000) Class A The instrument is Class A by product category.
Immunity
EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test meth-
ods, limits and performance achieved were:
a) EN61000-4-2 (1995) Electrostatic Discharge: 4 kV air, 4 kV contact Performance A.
b) EN61000-4-3 (1997) Electromagnetic Field: 3 V/m, 80 % AM at 1 kHz Performance A.
c) EN61000-4-11 (1994) Voltage Interrupt: 1 cycle, 100 % Performance A.
d)
EN61000-4-4 (1995) Fast Transient: 1 kV peak (ac line), 0.5 kV peak
(signal lines and RS232/GPIB ports)
Performance A.