November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names a
1551A Ex/1552A Ex Calibration Manual 2 Safety Information A Warning identifies conditions and actions that pose hazard(s) to the user. A Caution iden
Stik Thermometer Safety Information 3 If the Product overheats or is exposed to sudden physical shock, examine it for damage that can cause a safe
1551A Ex/1552A Ex Calibration Manual 4 Approvals The Product meets the above requirements using: EN 60079-0:2006 ELECTRICAL APPARATUS FOR EXPLOSIVE GA
Stik Thermometer Calibration 5 Nominal Probe Resistance at 0 °C ... 100 Ω Probe Hysteresis...
1551A Ex/1552A Ex Calibration Manual 6 Recommended Equipment The equipment listed in Table 2 will satisfy the calibration requirements: Table 2. Recom
Stik Thermometer Calibration 7 Record as-found Calibration Parameters Before any adjustments are made, the present values of the Product calibrati
1551A Ex/1552A Ex Calibration Manual 8 At each calibration point the measurement error may be calculated by subtracting the reference thermometer temp
Stik Thermometer Calibration 9 Adjustment at the Zero Calibration Point The next adjustment is made for the middle of the temperature range of the
1551A Ex/1552A Ex Calibration Manual 10 The command will set the value of the CAL:USER:TEMP3 parameter to the given temperature and automatically adju
Some semiconductors and custom IC's can bedamaged by electrostatic discharge duringhandling. This notice explains how you canminimize the chances
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and
5. USE STATIC SHIELDING CONTAINERS FOR HANDLING AND TRANSPORT.6. DO NOT SLIDE S.S. DEVICES OVER ANY SURFACE.7. AVOID PLASTIC,VINYL AND STYROFOAM
Stik Thermometer Maintenance 11 Maintenance Clean the Product WCaution To prevent possible damage to the product or to equipment under test, do no
1551A Ex/1552A Ex Calibration Manual 12 To change the batteries, see Figure 1: 1. Use a flat-blade screwdriver to remove the battery door screw. 2.
i Table of Contents Title Page Introduction...
1551A Ex/1552A Ex Calibration Manual ii
iii List of Tables Table Title Page 1. Symbols...
1551A Ex/1552A Ex Calibration Manual iv
v List of Figures Figure Title Page 1. Battery Replacement ...
1551A Ex/1552A Ex Calibration Manual vi
1 Introduction The Stik Thermometer (the Product) is a precision digital thermometer that can be used as an alternative to mercury liquid-in-glass (