® September 2006 ©2006 Fluke Corporation. All rights reserved. 1 114, 115, 116, and 117 Digital Multimeters Calibration Information Introduction XW
114, 115, 116, and 117 Calibration Information 10 Table 3. DMM Performance Tests (cont) Display Reading Step Function Range Applied 114[1] 115[1] 11
Digital Multimeters Performance Tests 11 Table 3. DMM Performance Tests (cont) Display Reading Step Function Range Applied 114[1] 115[1] 116[
114, 115, 116, and 117 Calibration Information 12 STBY78250+/-4..SCOPEOUTTRIGAUXA SENSE AUXVNORMALV RTDTRUE RMS MULTIMETER117VoltAlertCenter Meter t
Digital Multimeters Calibration Adjustment 13 2. Press g once to see the calibration counter. 3. Press g again to start the password entry.
114, 115, 116, and 117 Calibration Information 14 Meter Buttons Used in the Calibration Steps When performing the Calibration Adjustment Procedure, th
Digital Multimeters Calibration Adjustment 15 Notes Set the calibrator to Standby prior to changing the function switch position and after com
114, 115, 116, and 117 Calibration Information 16 Table 4. Calibration Adjustment Steps (cont) Calibration Steps Rotary Switch Position 114 115[1] 116
Digital Multimeters Replacement Parts 17 Replacement Parts Table 5 lists the Meter’s replacable parts identified in Figure 6. 6121023201721975
114, 115, 116, and 117 Calibration Information 18 Table 5. Replacable Parts List Item Description Part Number Qty. 1 LCD,FLUKE-11X,3.2V,TN,4-DIGIT,1/
Digital Multimeters Warranty 19 Table 5. Replaceable Parts List Item Description Part Number Qty. 19 BATTERY,PRIMARY,MNO2-ZN,9V,505MAH,6LR61,
114, 115, 116, and 117 Calibration Information 2 Safety Information "Warning" and "Caution" Statements A “XW Warning" ident
114, 115, 116, and 117 Calibration Information 20
Digital Multimeters International Electrical Symbols 3 International Electrical Symbols Table 1 lists the international symbols that appear in
114, 115, 116, and 117 Calibration Information 4 Accuracy Specifications Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model[1] DC
Digital Multimeters Basic Maintenance 5 Accuracy Specifications (cont) Function Range Resolution Accuracy ± ([% of Reading] + [Counts]) Model[
114, 115, 116, and 117 Calibration Information 6 HOLD MIN MA X RANGE10 AFUSEDCOMVALo / HOFFHzHzAUTO-VLoZVoltAlertAAVVmVVoltA le rtTRUE RMS MULTIMETE
Digital Multimeters Performance Tests 7 2. Remove the Meter from its holster. 3. Remove two screws from the case bottom. 4. Separate the ca
114, 115, 116, and 117 Calibration Information 8 Table 2. Required Equipment (cont) Recommended Equipment Measurement Function Accuracy 5500A Multi-
Digital Multimeters Performance Tests 9 2. Warm up the calibrator as required by its specifications. 3. Allow the temperature of the UUT to