80T-150UA
Universal Temperature Probe
Instruction Sheet
80T-150U
TEMPERATURE PROBE
C
F
OFF
Introduction
The Model 80T-150UA Universal Temperature Probe is a self-contained
temperature-to-voltage converter. The probe is designed to provide a
direct temperature reading when it is connected to any high impedance
DMM that is capable of 1 mV resolution and at least a 300-count full
scale readout capability. Output is 1 mV per degree (Celsius or
Fahrenheit). Two switch-selected temperature output scalings are
provided: -50 to +150 °C or -58 to +302 °F. The probe will stand off 60 V
dc or 30 V ac rms (42.4 V peak).
The unit is housed in two separate but attached assemblies: a
temperature probe and a temperature-to-voltage converter. The probe
contains the temperature-sensing element and is electrically connected to
the temperature-to-voltage converter through a 46-inch cable. A three-
position switch on the converter acts as a power switch and is used for
selecting Celsius or Fahrenheit scaling for the output. Two banana plugs
with standard 0.75-inch spacing are provided for connecting the 80T-
150UA to the DMM.
Operating power for the 80T-150UA is derived from a standard 9 V
battery. Typically, an alkaline battery provides more than 600 hours of
continuous operation before replacement is necessary. An OFF switch is
provided on the temperature-to-voltage converter to allow battery
conservation when the unit is not in use. In addition, the OFF position of
the power switch allows the battery condition to be determined via the
external DMM.
Temperature is measured by exposing the probe tip directly to the
material to be measured (non-corrosive liquid, gas, or solid). A direct
temperature reading is displayed on the DMM.
Symbols
XWarning: Risk of electric shock. Refer to the manual for
information.
W Attention: Refer to the manual for information about this
feature.
PN 3208738
February 2008
© 2008 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.
Vista de pagina 0
1 2 3 4 5 6

Indice de contenidos

Pagina 1 - 80T-150UA

® 80T-150UA Universal Temperature Probe Instruction Sheet 80T-150U TEMPERATURE PROBECFOFF Introduction The Model 80T-150UA Universal Temperature Probe

Pagina 2 - Specifications

Specifications The 80T-150UA will achieve rated accuracy when it is used with any 0.25 % DMM that has an input impedance of ≥1 MΩ. Electrical Measure

Pagina 3 - Operating Notes

General Weight: 150 g (5.3 oz) Overall Length: 53.8 inch (1.36 m) Battery: Standard 9 V battery (NEDA #1604,6F22,006P) Battery Life: 600+ hours, typic

Pagina 4 - General Maintenance

F1.eps Figure 1. Initial Case Temperature Above Ambient vs Meter Reading Above Ambient Operation Use the following procedure to operate the 80T-150UA

Pagina 5 - Cleaning

Battery Replacement XWWarning To avoid electrical shock, remove the probe from the measurement surface before opening the case. Totally reassemble the

Pagina 6 - Test Equipment Requirements

MP101 MP102Sensing Probe and Cable AssemblyVR1VR3VR4VR2 F4.eps Figure 2. 80T-150UA Temperature Probe Test Equipment Requirements Instrument Type Min

Comentarios a estos manuales

Sin comentarios