®PN 1554518 January 2000 ©2000 Fluke Corporation. All rights reserved. Printed in U.S.A. 1Model 65Infrared ThermometerCalibration InformationIntrodu
Model 65Calibration Information2Safety InformationA WarningTo avoid possible personal injury:• The information provided in this manual is for the use
Specifications3SpecificationsMeasurement Range −40 °C to +500 °C (−40 °F to +932 °F)Operating Temperature (Ambient) 0 °C to +50 °C (+32 °F to +122 °F)
Model 65Calibration Information4MaintenanceInstalling the BatteriesRemove the holster to access the battery door. Figure 1 shows how to replace the ba
Required Equipment5Required EquipmentTable 1 shows the equipment required for the performance tests and calibration adjustments.Table 1. Required Equi
Model 65Calibration Information6Table 2. Performance Test ValuesTemperature at 0.95 Emissivity UUT Reading Limits1, 2Room temperature Room temperature
Calibration Adjustments7MINMAX°C°FRoom temperaturecalibration modeHigh temperaturecalibration modeEnd calibrationmodePress to exitcalibration mode
Model 65Calibration Information88. Press D to enter the high-temperature adjustment mode (display shows 2_H), as shown in Figure 2.9. Press G. The d